We develop a wave-based tomographic imaging algorithm based upon a single rotating radially outward oriented transducer. At successive angular locations at a fixed radius, the transducer launches a primary field and collects the backscattered field in a ''pitch/catch'' operation. The hardware configuration, operating mode, and data collection method is identical to that of most medical intravascular ultrasound (IVUS) systems. IVUS systems form images of the medium surrounding the probe based upon ultrasonic B-scans, using a straight-ray model of sound propagation. Our goal is to develop a wave-based imaging algorithm using diffraction tomography techniques. Given the hardware configuration and the imaging …
continued below
Publisher Info:
Lawrence Livermore National Lab., Livermore, CA (United States)
Place of Publication:
Livermore, California
Provided By
UNT Libraries Government Documents Department
Serving as both a federal and a state depository library, the UNT Libraries Government Documents Department maintains millions of items in a variety of formats. The department is a member of the FDLP Content Partnerships Program and an Affiliated Archive of the National Archives.
Descriptive information to help identify this report.
Follow the links below to find similar items on the Digital Library.
Description
We develop a wave-based tomographic imaging algorithm based upon a single rotating radially outward oriented transducer. At successive angular locations at a fixed radius, the transducer launches a primary field and collects the backscattered field in a ''pitch/catch'' operation. The hardware configuration, operating mode, and data collection method is identical to that of most medical intravascular ultrasound (IVUS) systems. IVUS systems form images of the medium surrounding the probe based upon ultrasonic B-scans, using a straight-ray model of sound propagation. Our goal is to develop a wave-based imaging algorithm using diffraction tomography techniques. Given the hardware configuration and the imaging method, we refer to this system as ''radial reflection diffraction tomography.'' We consider two hardware configurations: a multimonostatic mode using a single transducer as described above, and a multistatic mode consisting of a single transmitter and an aperture formed by multiple receivers. In this latter case, the entire source/receiver aperture rotates about the fixed radius. Practically, such a probe is mounted at the end of a catheter or snaking tube that can be inserted into a part or medium with the goal of forming images of the plane perpendicular to the axis of rotation. We derive an analytic expression for the multimonostatic inverse but ultimately use the new Hilbert space inverse wave (HSIW) algorithm to construct images using both operating modes. Applications include improved IVUS imaging, bore hole tomography, and non-destructive evaluation (NDE) of parts with existing access holes.
This report is part of the following collection of related materials.
Office of Scientific & Technical Information Technical Reports
Reports, articles and other documents harvested from the Office of Scientific and Technical Information.
Office of Scientific and Technical Information (OSTI) is the Department of Energy (DOE) office that collects, preserves, and disseminates DOE-sponsored research and development (R&D) results that are the outcomes of R&D projects or other funded activities at DOE labs and facilities nationwide and grantees at universities and other institutions.
Lehman, S. K. & Norton, S. J.Radial Reflection Diffraction Tomography,