Search Results

open access

Errors in skin temperature measurements.

Description: Numerical simulation is used to investigate the accuracy of a direct-contact device for measuring skin-surface temperature. A variation of thermal conductivity of the foam has greater effect on the error rather than a variation of the blood perfusion rate. For a thermal conductivity of zero, an error of 1.5 oC in temperature was identified. For foam pad conductivities of 0.03 and 0.06 W/m-oC, the errors are 0.5 and 0.15 oC. For the transient study, with k=0 W/m-oC, it takes 4,900 seconds fo… more
Date: December 2008
Creator: Dugay, Murielle
Partner: UNT Libraries
Back to Top of Screen