CdWO4-Boron FY 2000 Task 4 Completion Report
Description:
The fabrication of boron-covered crystal scintillation detectors is described. Bulk boron-loaded epoxy material was cast and cut into 0.5 mm-thick wafers that were mounted on CdWO{sub 4} and CsI(Tl) crystals. The crystals were mounted on miniature photomultiplier tubes and gamma spectra were obtained with the detectors. The ability of these small detectors to produce spectra that can be analyzed to provide isotopic identification has been demonstrated. In addition, the detector can produce a si…