Ion Chromatography of Soluble Cr(III) and Cr(VI)
Description:
Ion chromatography coupled with a conductivity detector was used to investigate the analysis of Cr(III) and Cr(VI) in aqueous samples. An IC methodology for Cr(III) was developed using a cation column and an eluent containing tartaric acid, ethylenediamine, and acetonitrile at pH 2.9. The detection limit of this method can reach 0.1 ppm level with good precision. Several operational parameters were evaluated during the regular use of the method. Comparison of the IC method with AA method showed…
more
Date:
August 1988
Creator:
Huang, Julie Shiong-Jiun