Pipe Cleaning Operating Procedures

PDF Version Also Available for Download.

Description

This cleaning procedure outlines the steps involved in cleaning the high purity argon lines associated with the DO calorimeters. The procedure is broken down into 7 cycles: system setup, initial flush, wash, first rinse, second rinse, final rinse and drying. The system setup involves preparing the pump cart, line to be cleaned, distilled water, and interconnecting hoses and fittings. The initial flush is an off-line flush of the pump cart and its plumbing in order to preclude contaminating the line. The wash cycle circulates the detergent solution (Micro) at 180 degrees Fahrenheit through the line to be cleaned. The first … continued below

Physical Description

11 pages

Creation Information

Clark, D. & Wu, J. January 24, 1991.

Context

This report is part of the collection entitled: Office of Scientific & Technical Information Technical Reports and was provided by the UNT Libraries Government Documents Department to the UNT Digital Library, a digital repository hosted by the UNT Libraries. It has been viewed 25 times. More information about this report can be viewed below.

Who

People and organizations associated with either the creation of this report or its content.

Authors

Publisher

Provided By

UNT Libraries Government Documents Department

Serving as both a federal and a state depository library, the UNT Libraries Government Documents Department maintains millions of items in a variety of formats. The department is a member of the FDLP Content Partnerships Program and an Affiliated Archive of the National Archives.

Contact Us

What

Descriptive information to help identify this report. Follow the links below to find similar items on the Digital Library.

Description

This cleaning procedure outlines the steps involved in cleaning the high purity argon lines associated with the DO calorimeters. The procedure is broken down into 7 cycles: system setup, initial flush, wash, first rinse, second rinse, final rinse and drying. The system setup involves preparing the pump cart, line to be cleaned, distilled water, and interconnecting hoses and fittings. The initial flush is an off-line flush of the pump cart and its plumbing in order to preclude contaminating the line. The wash cycle circulates the detergent solution (Micro) at 180 degrees Fahrenheit through the line to be cleaned. The first rinse is then intended to rid the line of the majority of detergent and only needs to run for 30 minutes and at ambient temperature. The second rinse (if necessary) should eliminate the remaining soap residue. The final rinse is then intended to be a check that there is no remaining soap or other foreign particles in the line, particularly metal 'chips.' The final rinse should be run at 180 degrees Fahrenheit for at least 90 minutes. The filters should be changed after each cycle, paying particular attention to the wash cycle and the final rinse cycle return filters. These filters, which should be bagged and labeled, prove that the pipeline is clean. Only distilled water should be used for all cycles, especially rinsing. The level in the tank need not be excessive, merely enough to cover the heater float switch. The final rinse, however, may require a full 50 gallons. Note that most of the details of the procedure are included in the initial flush description. This section should be referred to if problems arise in the wash or rinse cycles.

Physical Description

11 pages

Language

Item Type

Identifier

Unique identifying numbers for this report in the Digital Library or other systems.

Collections

This report is part of the following collection of related materials.

Office of Scientific & Technical Information Technical Reports

Reports, articles and other documents harvested from the Office of Scientific and Technical Information.

Office of Scientific and Technical Information (OSTI) is the Department of Energy (DOE) office that collects, preserves, and disseminates DOE-sponsored research and development (R&D) results that are the outcomes of R&D projects or other funded activities at DOE labs and facilities nationwide and grantees at universities and other institutions.

What responsibilities do I have when using this report?

When

Dates and time periods associated with this report.

Creation Date

  • January 24, 1991

Added to The UNT Digital Library

  • May 19, 2016, 3:16 p.m.

Description Last Updated

  • Sept. 26, 2017, 6:37 p.m.

Usage Statistics

When was this report last used?

Yesterday: 0
Past 30 days: 0
Total Uses: 25

Interact With This Report

Here are some suggestions for what to do next.

Start Reading

PDF Version Also Available for Download.

International Image Interoperability Framework

IIF Logo

We support the IIIF Presentation API

Clark, D. & Wu, J. Pipe Cleaning Operating Procedures, report, January 24, 1991; Batavia, Illinois. (https://digital.library.unt.edu/ark:/67531/metadc840258/: accessed June 16, 2024), University of North Texas Libraries, UNT Digital Library, https://digital.library.unt.edu; crediting UNT Libraries Government Documents Department.

Back to Top of Screen