Calibration of energy dispersive x-ray spectrometers for analysis of thin environmental samples

Description: Four separate techniques for calibrating energy dispersive x-ray spectrometers are described. They include the use of (1) individual evaporated elemental thin-film standards, (2) nebulized multielement standard solution deposits to determine relative elemental sensitivity factors, (3) a semi-empirical approach to calculate relative elemental sensitivity factors, and (4) thick pure element disks. The first three techniques are applicable for a broad range of elements. The utilization of nebulize… more
Date: January 26, 1976
Creator: Giauque, R. D.; Garrett, R. B. & Goda, L. Y.
Partner: UNT Libraries Government Documents Department