Technical assessment of Navitar Zoom 6000 optic and Sony HDC-X310 camera for MEMS presentations and training.

PDF Version Also Available for Download.

Description

This report evaluates a newly-available, high-definition, video camera coupled with a zoom optical system for microscopic imaging of micro-electro-mechanical systems. We did this work to support configuration of three document-camera-like stations as part of an installation in a new Microsystems building at Sandia National Laboratories. The video display walls to be installed as part of these three presentation and training stations are of extraordinary resolution and quality. The new availability of a reasonably-priced, cinema-quality, high-definition video camera offers the prospect of filling these displays with full-motion imaging of Sandia's microscopic products at a quality substantially beyond the quality of typical … continued below

Physical Description

23 p.

Creation Information

Diegert, Carl F. February 1, 2006.

Context

This report is part of the collection entitled: Office of Scientific & Technical Information Technical Reports and was provided by the UNT Libraries Government Documents Department to the UNT Digital Library, a digital repository hosted by the UNT Libraries. It has been viewed 68 times. More information about this report can be viewed below.

Who

People and organizations associated with either the creation of this report or its content.

Publisher

Provided By

UNT Libraries Government Documents Department

Serving as both a federal and a state depository library, the UNT Libraries Government Documents Department maintains millions of items in a variety of formats. The department is a member of the FDLP Content Partnerships Program and an Affiliated Archive of the National Archives.

Contact Us

What

Descriptive information to help identify this report. Follow the links below to find similar items on the Digital Library.

Description

This report evaluates a newly-available, high-definition, video camera coupled with a zoom optical system for microscopic imaging of micro-electro-mechanical systems. We did this work to support configuration of three document-camera-like stations as part of an installation in a new Microsystems building at Sandia National Laboratories. The video display walls to be installed as part of these three presentation and training stations are of extraordinary resolution and quality. The new availability of a reasonably-priced, cinema-quality, high-definition video camera offers the prospect of filling these displays with full-motion imaging of Sandia's microscopic products at a quality substantially beyond the quality of typical video microscopes. Simple and robust operation of the microscope stations will allow the extraordinary-quality imaging to contribute to Sandia's day-to-day research and training operations. This report illustrates the disappointing image quality from a camera/lens system comprised of a Sony HDC-X310 high-definition video camera coupled to a Navitar Zoom 6000 lens. We determined that this Sony camera is capable of substantially more image quality than the Navitar optic can deliver. We identified an optical doubler lens from Navitar as the component of their optical system that accounts for a substantial part of the image quality problem. While work continues to incrementally improve performance of the Navitar system, we are also evaluating optical systems from other vendors to couple to this Sony camera.

Physical Description

23 p.

Language

Item Type

Identifier

Unique identifying numbers for this report in the Digital Library or other systems.

Collections

This report is part of the following collection of related materials.

Office of Scientific & Technical Information Technical Reports

Reports, articles and other documents harvested from the Office of Scientific and Technical Information.

Office of Scientific and Technical Information (OSTI) is the Department of Energy (DOE) office that collects, preserves, and disseminates DOE-sponsored research and development (R&D) results that are the outcomes of R&D projects or other funded activities at DOE labs and facilities nationwide and grantees at universities and other institutions.

What responsibilities do I have when using this report?

When

Dates and time periods associated with this report.

Creation Date

  • February 1, 2006

Added to The UNT Digital Library

  • Sept. 21, 2016, 2:29 a.m.

Description Last Updated

  • Dec. 7, 2016, 10:58 p.m.

Usage Statistics

When was this report last used?

Yesterday: 0
Past 30 days: 0
Total Uses: 68

Interact With This Report

Here are some suggestions for what to do next.

Start Reading

PDF Version Also Available for Download.

International Image Interoperability Framework

IIF Logo

We support the IIIF Presentation API

Diegert, Carl F. Technical assessment of Navitar Zoom 6000 optic and Sony HDC-X310 camera for MEMS presentations and training., report, February 1, 2006; United States. (https://digital.library.unt.edu/ark:/67531/metadc876177/: accessed April 25, 2024), University of North Texas Libraries, UNT Digital Library, https://digital.library.unt.edu; crediting UNT Libraries Government Documents Department.

Back to Top of Screen