Description

An adiabatic surface thermometer was developed to control automatically the critical temperature-time cycle of a production vacuum-brazing process. Investigations revealed that optimum braze-joint strength required precise control of the brazing temperature. Spot-welded thermocouples could not be used because the spot welds cause surface damage. This thermometer touches the surface and uses a differential thermocouple and heater to measure surface temperature without heat flow, thereby eliminating large errors caused by conduction losses common to conventional spring-loaded thermocouples. Temperatures in air or vacuum are measured to 800$sup 0$C with errors less than 5$sup 0$C. This thermometer has minimized the rejection of production ... continued below

Physical Description

Pages: 13

Creation Information

Dittbenner, G.R. August 27, 1975.

Context

This article is part of the collection entitled: Office of Scientific & Technical Information Technical Reports and was provided by UNT Libraries Government Documents Department to Digital Library, a digital repository hosted by the UNT Libraries. More information about this article can be viewed below.

Who

People and organizations associated with either the creation of this article or its content.

Publisher

Provided By

UNT Libraries Government Documents Department

Serving as both a federal and a state depository library, the UNT Libraries Government Documents Department maintains millions of items in a variety of formats. The department is a member of the FDLP Content Partnerships Program and an Affiliated Archive of the National Archives.

Contact Us

What

Descriptive information to help identify this article. Follow the links below to find similar items on the Digital Library.

Description

An adiabatic surface thermometer was developed to control automatically the critical temperature-time cycle of a production vacuum-brazing process. Investigations revealed that optimum braze-joint strength required precise control of the brazing temperature. Spot-welded thermocouples could not be used because the spot welds cause surface damage. This thermometer touches the surface and uses a differential thermocouple and heater to measure surface temperature without heat flow, thereby eliminating large errors caused by conduction losses common to conventional spring-loaded thermocouples. Temperatures in air or vacuum are measured to 800$sup 0$C with errors less than 5$sup 0$C. This thermometer has minimized the rejection of production parts, resulting in a cost saving to the U. S. Energy Research and Development Administration. (auth)

Physical Description

Pages: 13

Notes

Dep. NTIS

Source

  • SME/ERDA welding and joining workshop, Chicago, Illinois, USA, 28 Oct 1975

Language

Item Type

Identifier

Unique identifying numbers for this article in the Digital Library or other systems.

  • Report No.: UCRL--77295
  • Report No.: CONF-751012--8
  • Grant Number: None
  • Office of Scientific & Technical Information Report Number: 4167006
  • Archival Resource Key: ark:/67531/metadc870228

Collections

This article is part of the following collection of related materials.

Office of Scientific & Technical Information Technical Reports

What responsibilities do I have when using this article?

When

Dates and time periods associated with this article.

Creation Date

  • August 27, 1975

Added to The UNT Digital Library

  • Sept. 16, 2016, 12:32 a.m.

Description Last Updated

  • Oct. 18, 2016, 1:07 p.m.

Usage Statistics

When was this article last used?

Congratulations! It looks like you are the first person to view this item online.

Interact With This Article

Here are some suggestions for what to do next.

Start Reading

PDF Version Also Available for Download.

Citations, Rights, Re-Use

Dittbenner, G.R. Adiabatic surface thermometer for improved production braze quality, article, August 27, 1975; Livermore, California. (digital.library.unt.edu/ark:/67531/metadc870228/: accessed August 23, 2017), University of North Texas Libraries, Digital Library, digital.library.unt.edu; crediting UNT Libraries Government Documents Department.