Procedures for the routine calibration of gamma-ray scintillation spectrometers and multichannel pulse-height analyzers are presented. A scintillation spectrometer, designed for use of recently developed computer data processing techniques, is described. The performance specifications required of all electronic components in this application, particularly the performance specifications for multi-channel pulse-height analyzers, are treated; and simple tests are described for determining the performance specifications of an analyzer. (auth)
Publisher Info:
Phillips Petroleum Co. Atomic Energy Div., Idaho Falls, Idaho
Place of Publication:
Idaho Falls, Idaho
Provided By
UNT Libraries Government Documents Department
Serving as both a federal and a state depository library, the UNT Libraries Government Documents Department maintains millions of items in a variety of formats. The department is a member of the FDLP Content Partnerships Program and an Affiliated Archive of the National Archives.
Descriptive information to help identify this report.
Follow the links below to find similar items on the Digital Library.
Description
Procedures for the routine calibration of gamma-ray scintillation spectrometers and multichannel pulse-height analyzers are presented. A scintillation spectrometer, designed for use of recently developed computer data processing techniques, is described. The performance specifications required of all electronic components in this application, particularly the performance specifications for multi-channel pulse-height analyzers, are treated; and simple tests are described for determining the performance specifications of an analyzer. (auth)
This report is part of the following collection of related materials.
Office of Scientific & Technical Information Technical Reports
Reports, articles and other documents harvested from the Office of Scientific and Technical Information.
Office of Scientific and Technical Information (OSTI) is the Department of Energy (DOE) office that collects, preserves, and disseminates DOE-sponsored research and development (R&D) results that are the outcomes of R&D projects or other funded activities at DOE labs and facilities nationwide and grantees at universities and other institutions.
Crouch, D.F. & Heath, R.L.ROUTINE TESTING AND CALIBRATION PROCEDURES FOR MULTICHANNEL PULSE ANALYZERS AND GAMMA-RAY SPECTROMETERS,