A practical, prototype silicon p-n junction fast-neutron dosimeter, sensitive in the same range as human tissue, was developed, together with sn associated read-out circuit to facilitate the accurate measurement of accumulated dose. From both theoretical and experimental considerations, it was demonstrated that the dosimeter is essentially insensitive to the gamma and thermal components of a uranium fission spectrum. It was shown that accumulated damage effects appear to be environmentally stable up to an ambient temperature of 100 C. A rather raarked reversible temperature dependence of the read-out parameters requires either control of the read-out temperature or temperature compensation in the …
continued below
Publisher Info:
Battelle Memorial Inst., Columbus, Ohio
Place of Publication:
Columbus, Ohio
Provided By
UNT Libraries Government Documents Department
Serving as both a federal and a state depository library, the UNT Libraries Government Documents Department maintains millions of items in a variety of formats. The department is a member of the FDLP Content Partnerships Program and an Affiliated Archive of the National Archives.
Descriptive information to help identify this report.
Follow the links below to find similar items on the Digital Library.
Description
A practical, prototype silicon p-n junction fast-neutron dosimeter, sensitive in the same range as human tissue, was developed, together with sn associated read-out circuit to facilitate the accurate measurement of accumulated dose. From both theoretical and experimental considerations, it was demonstrated that the dosimeter is essentially insensitive to the gamma and thermal components of a uranium fission spectrum. It was shown that accumulated damage effects appear to be environmentally stable up to an ambient temperature of 100 C. A rather raarked reversible temperature dependence of the read-out parameters requires either control of the read-out temperature or temperature compensation in the read-out device. A high degree of reproducibility of dosimeter characteristics from one device to another was not achieved. The lack of reproducibility was attributed to uncontrolled variables in the bulk silicon from which the devices are fabricated, and in the production procedure. (auth)
This report is part of the following collection of related materials.
Office of Scientific & Technical Information Technical Reports
Reports, articles and other documents harvested from the Office of Scientific and Technical Information.
Office of Scientific and Technical Information (OSTI) is the Department of Energy (DOE) office that collects, preserves, and disseminates DOE-sponsored research and development (R&D) results that are the outcomes of R&D projects or other funded activities at DOE labs and facilities nationwide and grantees at universities and other institutions.
Gorton, H. C.; Mengali, O. J.; Zacaroli, A. R.; Crooks, R. K.; Swartz, J. M. & Peet, C. S.Experimental and Research Work in Neutron Dosimetry. Final Summary Report for the Period May 15, 1959-June 15, 1960,