In this report, data from neutron measurement and dosimetry studies performed by Pacific Northwest Laboratory, the Environmental Measurements Laboratory, and Rensselaer Polytechnic Institute are examined and compared. The purpose of this data correlation effort is to determine whether useful relationships exist between the actual neutron dose equivalent in a typical commercial nuclear power reactor and various measurement parameters, such as ratios of the response of 9-in. to 3-in. spheres, neutron/gamma ratios, albedo dosimeter response and neutron spectrometer readings. In most neutron radiation fields found in the reactors visited, the response of albedo dosimeters can be brought into reasonable agreement with …
continued below
Publisher Info:
Pacific Northwest National Laboratory (PNNL), Richland, WA (United States)
Place of Publication:
Richland, Washington
Provided By
UNT Libraries Government Documents Department
Serving as both a federal and a state depository library, the UNT Libraries Government Documents Department maintains millions of items in a variety of formats. The department is a member of the FDLP Content Partnerships Program and an Affiliated Archive of the National Archives.
Descriptive information to help identify this report.
Follow the links below to find similar items on the Digital Library.
Description
In this report, data from neutron measurement and dosimetry studies performed by Pacific Northwest Laboratory, the Environmental Measurements Laboratory, and Rensselaer Polytechnic Institute are examined and compared. The purpose of this data correlation effort is to determine whether useful relationships exist between the actual neutron dose equivalent in a typical commercial nuclear power reactor and various measurement parameters, such as ratios of the response of 9-in. to 3-in. spheres, neutron/gamma ratios, albedo dosimeter response and neutron spectrometer readings. In most neutron radiation fields found in the reactors visited, the response of albedo dosimeters can be brought into reasonable agreement with dose equivalents measured with multispheres, tissue equivalent proportional counters (TEPCs) or remmeters. Because the responses of the remmeters, like the responses of albedo dosimeters, are energy dependent, it is preferable to correct the responses of the albedo dosimeters to agree with dose equivalents measured with either TEPCs or multispheres. If one of these laboratory systems has been used to measure neutron dose equivalents at a specific pressurized water reactor, a calculated average albedo dosimeter correction factor can be used for most locations at that reactor. However, if the measured 9-in. to 3-in. remmeter ratio is greater than 0.20, it is advisable to use a plot of 9-in. to 3-in. remmeter ratios versus albedo dosimeter correction factors to obtain an albedo dosimeter correction factor. Because 9-in. to 3-in. remmeter ratios at boiling water reactors are typically greater than 0.20, the latter approach applies to this type of reactor.
This report is part of the following collection of related materials.
Office of Scientific & Technical Information Technical Reports
Reports, articles and other documents harvested from the Office of Scientific and Technical Information.
Office of Scientific and Technical Information (OSTI) is the Department of Energy (DOE) office that collects, preserves, and disseminates DOE-sponsored research and development (R&D) results that are the outcomes of R&D projects or other funded activities at DOE labs and facilities nationwide and grantees at universities and other institutions.
Rathbun, L. A. & Endres, G. W.R.Correlation of Neutron Data Taken at Commercial Nuclear Sites,
report,
October 1, 1983;
Richland, Washington.
(https://digital.library.unt.edu/ark:/67531/metadc831363/:
accessed March 19, 2024),
University of North Texas Libraries, UNT Digital Library, https://digital.library.unt.edu;
crediting UNT Libraries Government Documents Department.