This paper presents an alternative method for verifying electronic balance linearity and accuracy. This method is being developed for safeguards weighings (weighings for the control and accountability of nuclear material) at the Idaho National Engineering and Environmental Laboratory (INEEL). With regard to balance linearity and accuracy, DOE Order 5633.3B, Control and Accountability of Nuclear Materials, Paragraph 2, 4, e, (1), (a) Scales and Balances Program, states: ``All scales and balances used for accountability purposes shall be maintained in good working condition, recalibrated according to an established schedule, and checked for accuracy and linearity on each day that the scale or balance is used for accountability purposes.`` Various tests have been proposed for testing accuracy and linearity. In the 1991 Measurement Science Conference, Dr. Walter E. Kupper presented a paper entitled: ``Validation of High Accuracy Weighing Equipment.`` Dr. Kupper emphasized that tolerance checks for calibrated, state-of-the-art electronic equipment need not be complicated, and he presented four easy steps for verifying that a calibrated balance is operating correctly. These tests evaluate the standard deviation of successive weighings (of the same load), the off-center error, the calibration error, and the error due to nonlinearity. This method of balance validation is undoubtedly an authoritative means of ensuring balance operability, yet it could have two drawbacks: one, the test for linearity is not intuitively obvious, especially from a statistical viewpoint; and two, there is an absence of definitively defined testing limits. Hence, this paper describes an alternative means of verifying electronic balance linearity and accuracy that is being developed for safeguards measurements at the INEEL.