This paper presents an alternative method for verifying electronic balance linearity and accuracy. This method is being developed for safeguards weighings (weighings for the control and accountability of nuclear material) at the Idaho National Engineering and Environmental Laboratory (INEEL). With regard to balance linearity and accuracy, DOE Order 5633.3B, Control and Accountability of Nuclear Materials, Paragraph 2, 4, e, (1), (a) Scales and Balances Program, states: ``All scales and balances used for accountability purposes shall be maintained in good working condition, recalibrated according to an established schedule, and checked for accuracy and linearity on each day that the scale or ... continued below

Physical Description

16 p.

Creation Information

Thomas, I.R. February 1, 1998.

Context

This article is part of the collection entitled: Office of Scientific & Technical Information Technical Reports and was provided by UNT Libraries Government Documents Department to Digital Library, a digital repository hosted by the UNT Libraries. It has been viewed 40 times . More information about this article can be viewed below.

Who

People and organizations associated with either the creation of this article or its content.

Author

Sponsor

Publisher

Provided By

UNT Libraries Government Documents Department

Serving as both a federal and a state depository library, the UNT Libraries Government Documents Department maintains millions of items in a variety of formats. The department is a member of the FDLP Content Partnerships Program and an Affiliated Archive of the National Archives.

Contact Us

What

Descriptive information to help identify this article. Follow the links below to find similar items on the Digital Library.

Description

This paper presents an alternative method for verifying electronic balance linearity and accuracy. This method is being developed for safeguards weighings (weighings for the control and accountability of nuclear material) at the Idaho National Engineering and Environmental Laboratory (INEEL). With regard to balance linearity and accuracy, DOE Order 5633.3B, Control and Accountability of Nuclear Materials, Paragraph 2, 4, e, (1), (a) Scales and Balances Program, states: ``All scales and balances used for accountability purposes shall be maintained in good working condition, recalibrated according to an established schedule, and checked for accuracy and linearity on each day that the scale or balance is used for accountability purposes.`` Various tests have been proposed for testing accuracy and linearity. In the 1991 Measurement Science Conference, Dr. Walter E. Kupper presented a paper entitled: ``Validation of High Accuracy Weighing Equipment.`` Dr. Kupper emphasized that tolerance checks for calibrated, state-of-the-art electronic equipment need not be complicated, and he presented four easy steps for verifying that a calibrated balance is operating correctly. These tests evaluate the standard deviation of successive weighings (of the same load), the off-center error, the calibration error, and the error due to nonlinearity. This method of balance validation is undoubtedly an authoritative means of ensuring balance operability, yet it could have two drawbacks: one, the test for linearity is not intuitively obvious, especially from a statistical viewpoint; and two, there is an absence of definitively defined testing limits. Hence, this paper describes an alternative means of verifying electronic balance linearity and accuracy that is being developed for safeguards measurements at the INEEL.

Physical Description

16 p.

Notes

INIS; OSTI as DE98052757

Source

  • 1998 measurement science conference symposium, Pasadena, CA (United States), 5-6 Feb 1998

Language

Item Type

Identifier

Unique identifying numbers for this article in the Digital Library or other systems.

  • Other: DE98052757
  • Report No.: INEEL/CON--98-00055
  • Report No.: CONF-980230--
  • Grant Number: AC07-94ID13223
  • Office of Scientific & Technical Information Report Number: 653958
  • Archival Resource Key: ark:/67531/metadc703188

Collections

This article is part of the following collection of related materials.

Office of Scientific & Technical Information Technical Reports

Reports, articles and other documents harvested from the Office of Scientific and Technical Information.

Office of Scientific and Technical Information (OSTI) is the Department of Energy (DOE) office that collects, preserves, and disseminates DOE-sponsored research and development (R&D) results that are the outcomes of R&D projects or other funded activities at DOE labs and facilities nationwide and grantees at universities and other institutions.

What responsibilities do I have when using this article?

When

Dates and time periods associated with this article.

Creation Date

  • February 1, 1998

Added to The UNT Digital Library

  • Sept. 12, 2015, 6:31 a.m.

Description Last Updated

  • June 13, 2016, 6:37 p.m.

Usage Statistics

When was this article last used?

Yesterday: 0
Past 30 days: 0
Total Uses: 40

Interact With This Article

Here are some suggestions for what to do next.

Start Reading

PDF Version Also Available for Download.

International Image Interoperability Framework

IIF Logo

We support the IIIF Presentation API

Thomas, I.R. An alternative test for verifying electronic balance linearity, article, February 1, 1998; Idaho Falls, Idaho. (digital.library.unt.edu/ark:/67531/metadc703188/: accessed December 12, 2018), University of North Texas Libraries, Digital Library, digital.library.unt.edu; crediting UNT Libraries Government Documents Department.