Description

Los Alamos National Laboratories has completed the design, manufacture and calibration of a vacuum-compatible, tungsten lamp, integrated sphere. The light source has been calibrated at the National Institute of Standards and Technology (NIST) and is intended for use as a calibration standard for remote sensing instrumentation. Calibration 2{sigma} uncertainty varied with wavelength from 1.21% at 400 nm and 0.73% at 900 nm, to 3.95% at 2,400 nm. The inner radius of the Spectralon-coated sphere is 21.2 cm with a 7.4 cm square exit aperture. A small satellite sphere is attached to the main sphere and its output coupled through a ... continued below

Physical Description

13 p.

Creation Information

Byrd, D.A.; Atkins, W.H.; Bender, S.C.; Christensen, R.W. & Michaud, F.D. March 1, 1999.

Context

This article is part of the collection entitled: Office of Scientific & Technical Information Technical Reports and was provided by UNT Libraries Government Documents Department to Digital Library, a digital repository hosted by the UNT Libraries. More information about this article can be viewed below.

Who

People and organizations associated with either the creation of this article or its content.

Sponsor

Publisher

Provided By

UNT Libraries Government Documents Department

Serving as both a federal and a state depository library, the UNT Libraries Government Documents Department maintains millions of items in a variety of formats. The department is a member of the FDLP Content Partnerships Program and an Affiliated Archive of the National Archives.

Contact Us

What

Descriptive information to help identify this article. Follow the links below to find similar items on the Digital Library.

Description

Los Alamos National Laboratories has completed the design, manufacture and calibration of a vacuum-compatible, tungsten lamp, integrated sphere. The light source has been calibrated at the National Institute of Standards and Technology (NIST) and is intended for use as a calibration standard for remote sensing instrumentation. Calibration 2{sigma} uncertainty varied with wavelength from 1.21% at 400 nm and 0.73% at 900 nm, to 3.95% at 2,400 nm. The inner radius of the Spectralon-coated sphere is 21.2 cm with a 7.4 cm square exit aperture. A small satellite sphere is attached to the main sphere and its output coupled through a stepper motor driven aperture. The variable aperture allows a constant radiance without effecting the color temperature output from the main sphere. The sphere`s output is transmitted into a vacuum test environment through a fused silica window that is an integral part of the outer housing of the vacuum shell assembly. The atmosphere within this outer housing is composed of 240 K nitrogen gas, provided by a custom LN{sub 2} vaporizer unit. Use of the nitrogen gas maintains the internal temperature of the sphere at a nominal 300 K {+-}10{degree}. The calibrated spectral range of the source is 0.4 {micro}m through 2.4 {micro}m. Three, color temperature matched, 20 W bulbs together with a 10 W bulb are within the main integrating sphere. Two 20 W bulbs, also color temperature matched, reside in the satellite integrating sphere. A Silicon and a Germanium broadband detector are situated within the inner surface of the main sphere. Their purpose is for the measurement of the internal broadband irradiance. A fiber-optic-coupled spectrometer measures the internal color temperature that is maintained by current control on the lamps. Each lamp is independently operated allowing for radiances with common color temperatures ranging from near 0.026 W/cm{sup 2}/sr to about 0.1 W/cm{sup 2}/sr at a wavelength of 0.9 {micro}m (the location of the peak spectral radiance).

Physical Description

13 p.

Notes

OSTI as DE99002082

Source

  • 43. international symposium on optical science, engineering, and instrumentation, San Diego, CA (United States), 19-24 Jul 1998

Language

Item Type

Identifier

Unique identifying numbers for this article in the Digital Library or other systems.

  • Other: DE99002082
  • Report No.: LA-UR--99-3144
  • Report No.: CONF-980731--
  • Grant Number: W-7405-ENG-36
  • Office of Scientific & Technical Information Report Number: 325780
  • Archival Resource Key: ark:/67531/metadc681263

Collections

This article is part of the following collection of related materials.

Office of Scientific & Technical Information Technical Reports

Reports, articles and other documents harvested from the Office of Scientific and Technical Information.

Office of Scientific and Technical Information (OSTI) is the Department of Energy (DOE) office that collects, preserves, and disseminates DOE-sponsored research and development (R&D) results that are the outcomes of R&D projects or other funded activities at DOE labs and facilities nationwide and grantees at universities and other institutions.

What responsibilities do I have when using this article?

When

Dates and time periods associated with this article.

Creation Date

  • March 1, 1999

Added to The UNT Digital Library

  • July 25, 2015, 2:20 a.m.

Description Last Updated

  • Feb. 26, 2016, 9:39 p.m.

Usage Statistics

When was this article last used?

Yesterday: 0
Past 30 days: 0
Total Uses: 2

Interact With This Article

Here are some suggestions for what to do next.

Start Reading

PDF Version Also Available for Download.

International Image Interoperability Framework

IIF Logo

We support the IIIF Presentation API

Byrd, D.A.; Atkins, W.H.; Bender, S.C.; Christensen, R.W. & Michaud, F.D. Vacuum-compatible standard diffuse source, manufacture and calibration, article, March 1, 1999; New Mexico. (digital.library.unt.edu/ark:/67531/metadc681263/: accessed August 19, 2018), University of North Texas Libraries, Digital Library, digital.library.unt.edu; crediting UNT Libraries Government Documents Department.