Sealed source and device design safety testing. Volume 5: Technical report on the findings of Task 4, Investigation of failed radioactive stainless steel troxler gauges

PDF Version Also Available for Download.

Description

This report covers the Task 4 activities for the Sealed Source and Device Safety testing program. SwRI was contracted to investigate failed radioactive stainless steel troxler gauges. SwRI`s task was to determine the cause of failure of the rods and the extent of the problem. SwRI concluded that the broken rod failed in a brittle manner due to a hard zone in the heat affected zone.

Physical Description

48 p.

Creation Information

Benac, D.J. & Schick, W.R. October 1, 1995.

Context

This report is part of the collection entitled: Office of Scientific & Technical Information Technical Reports and was provided by UNT Libraries Government Documents Department to Digital Library, a digital repository hosted by the UNT Libraries. More information about this report can be viewed below.

Who

People and organizations associated with either the creation of this report or its content.

Authors

Sponsor

Provided By

UNT Libraries Government Documents Department

Serving as both a federal and a state depository library, the UNT Libraries Government Documents Department maintains millions of items in a variety of formats. The department is a member of the FDLP Content Partnerships Program and an Affiliated Archive of the National Archives.

Contact Us

What

Descriptive information to help identify this report. Follow the links below to find similar items on the Digital Library.

Description

This report covers the Task 4 activities for the Sealed Source and Device Safety testing program. SwRI was contracted to investigate failed radioactive stainless steel troxler gauges. SwRI`s task was to determine the cause of failure of the rods and the extent of the problem. SwRI concluded that the broken rod failed in a brittle manner due to a hard zone in the heat affected zone.

Physical Description

48 p.

Notes

INIS; OSTI as TI96002833

Source

  • Other Information: PBD: Oct 1995

Language

Item Type

Identifier

Unique identifying numbers for this report in the Digital Library or other systems.

  • Other: TI96002833
  • Report No.: NUREG/CR--6074-Vol.5
  • DOI: 10.2172/137424 | External Link
  • Office of Scientific & Technical Information Report Number: 137424
  • Archival Resource Key: ark:/67531/metadc625621

Collections

This report is part of the following collection of related materials.

Office of Scientific & Technical Information Technical Reports

What responsibilities do I have when using this report?

When

Dates and time periods associated with this report.

Creation Date

  • October 1, 1995

Added to The UNT Digital Library

  • June 16, 2015, 7:43 a.m.

Description Last Updated

  • April 7, 2016, 8:37 p.m.

Usage Statistics

When was this report last used?

Yesterday: 0
Past 30 days: 0
Total Uses: 3

Interact With This Report

Here are some suggestions for what to do next.

Start Reading

PDF Version Also Available for Download.

Citations, Rights, Re-Use

Benac, D.J. & Schick, W.R. Sealed source and device design safety testing. Volume 5: Technical report on the findings of Task 4, Investigation of failed radioactive stainless steel troxler gauges, report, October 1, 1995; United States. (digital.library.unt.edu/ark:/67531/metadc625621/: accessed September 24, 2017), University of North Texas Libraries, Digital Library, digital.library.unt.edu; crediting UNT Libraries Government Documents Department.