An isochronal differential-type microcalorimeter has been designed and constructed. As a result of its simple design it is very easy to handle the samples and assemble the calorimeter. Important to the operation of the calorimeter is a program, also working on the differential principle, that provides linear temperature rise of the samples. This calorimeter is used to measure very small energy releases such as those found in precipitation, stored energy, etc. It is demonstrated that the calorimeter is easily capable of measuring 0.0005 cal with a probable error of the order of 1% to 2%.
We've identified this
report
as a primary source within our collections. Researchers, educators, and students may find this report useful in their work.
Provided By
UNT Libraries Government Documents Department
Serving as both a federal and a state depository library, the UNT Libraries Government Documents Department maintains millions of items in a variety of formats. The department is a member of the FDLP Content Partnerships Program and an Affiliated Archive of the National Archives.
Added Title:
Brookhaven National Laboratory Report BNL-6676
Description
An isochronal differential-type microcalorimeter has been designed and constructed. As a result of its simple design it is very easy to handle the samples and assemble the calorimeter. Important to the operation of the calorimeter is a program, also working on the differential principle, that provides linear temperature rise of the samples. This calorimeter is used to measure very small energy releases such as those found in precipitation, stored energy, etc. It is demonstrated that the calorimeter is easily capable of measuring 0.0005 cal with a probable error of the order of 1% to 2%.
This report is part of the following collections of related materials.
TRAIL Microcard Collection
Imaged from microcard, these technical reports describe research performed for U.S. government agencies from the 1930s to the 1960s. The reports were provided by the Technical Report Archive and Image Library (TRAIL).
The Technical Report Archive & Image Library (TRAIL) identifies, acquires, catalogs, digitizes and provides unrestricted access to U.S. government agency technical reports. The mission of TRAIL is to ensure preservation, discoverability, and persistent open access to government technical publications regardless of form or format.
Arndt, R. A. & Fujita, F. E.An Isochronal Differential Microcalorimeter,
report,
February 25, 1963;
Washington D.C..
(https://digital.library.unt.edu/ark:/67531/metadc1255819/:
accessed February 17, 2025),
University of North Texas Libraries, UNT Digital Library, https://digital.library.unt.edu;
crediting UNT Libraries Government Documents Department.