Some comments on manganin wire pressure gauges Page: 1 of 11
This report is part of the collection entitled: Office of Scientific & Technical Information Technical Reports and was provided to UNT Digital Library by the UNT Libraries Government Documents Department.
Extracted Text
The following text was automatically extracted from the image on this page using optical character recognition software:
Some Cocnments on Manganin Wire Pressure Gauges *
Carl Andeen and Donald Schuele
Case Western Reserve University, Cleveland., Ohio 44106
and
John Fontanella {- 77
United States Naval Academy., Annapolis, Ed, 21402A standard manganin wire pressure gauge is
examined bry comparing it with a recently developed
0.01% CaF2 capacitive pressure gauge. The effects
of the Bridgman ."water kick" and. intrinsic time
constant are clearly shown and the results
. punctuate the usual assertion that manganin wire
gauges are extremely difficult to use approachirg
the 0.1% level.NOTICE
This report was prepared as an account of work
sponsored by the United States Government. Neither
the United States nor the United States Atomic Energy
Commission, nor any of their employees, nor any of
their contractors, subcontractors, or their employees,
makes any warranty, express or implied, or assumes any
legal liability or responsibility for the accuracy, com-
pleteness or usefulness of any information, apparatus,
product or process disclosed, or represents that its use
would not infringe privately owned rights.-1--
M ASTE
DISTRIBUTION OF THIS DOCUMENT IS UNLIMIT. D
Q~o0 - (,i
t
Upcoming Pages
Here’s what’s next.
Search Inside
This report can be searched. Note: Results may vary based on the legibility of text within the document.
Tools / Downloads
Get a copy of this page or view the extracted text.
Citing and Sharing
Basic information for referencing this web page. We also provide extended guidance on usage rights, references, copying or embedding.
Reference the current page of this Report.
Andeen, C.; Schuele, D. & Fontanella, J. Some comments on manganin wire pressure gauges, report, January 1, 1973; Cleveland, Ohio. (https://digital.library.unt.edu/ark:/67531/metadc1023599/m1/1/: accessed April 19, 2024), University of North Texas Libraries, UNT Digital Library, https://digital.library.unt.edu; crediting UNT Libraries Government Documents Department.