Results from the Cooler and Lead Tests Metadata

Metadata describes a digital item, providing (if known) such information as creator, publisher, contents, size, relationship to other resources, and more. Metadata may also contain "preservation" components that help us to maintain the integrity of digital files over time.

Title

  • Main Title Results from the Cooler and Lead Tests

Creator

  • Author: Green, Michael A
    Creator Type: Personal

Contributor

  • Sponsor: Accelerator&amp
    Contributor Type: Organization
  • Sponsor: Fusion Research Division
    Contributor Type: Organization
  • Sponsor: Engineering Division
    Contributor Type: Organization

Publisher

  • Name: Lawrence Berkeley National Laboratory
    Place of Publication: Berkeley, California
    Additional Info: Ernest Orlando Lawrence Berkeley National Laboratory, Berkeley, CA (United States)

Date

  • Creation: 2010-06-10

Language

  • English

Description

  • Content Description: The report presents the results of testing MICE spectrometer magnet current leads on a test apparatus that combines both the copper leads and the high temperature superconducting (HTS) leads with a single Cryomech PT415 cooler and liquid helium tank. The current is carried through the copper leads from 300 K to the top of the HTS leads. The current is then carried through the HTS leads to a feed-through from the vacuum space to the inside of a liquid helium tank. The experiment allows one to measure the performance of both cooler stages along with the performance of the leads. While the leads were powered we measured the voltage drops through the copper leads, through the HTS leads, through spliced to the feed-through, through the feed-through and through the low-temperature superconducting loop that connects one lead to the other. Measurements were made using the leads that were used in spectrometer magnet 1A and spectrometer magnet 2A. These are the same leads that were used for Superbend and Venus magnets at LBNL. The IL/A for these leads was 5.2 x 10{sup 6} m{sup -1}. The leads turned out to be too long. The same measurements were made using the leads that were installed in magnet 2B. The magnet 2B leads had an IL/A of 3.3 x 10{sup 6} A m{sup -1}. This report discusses the cooler performance and the measured electrical performance of the lead circuit that contains the copper leads and the superconducting leads. All of the HTS leads that were installed in magnet 2B were current tested using this apparatus.
  • Physical Description: 42

Subject

  • Keyword: Copper
  • Keyword: Spectrometers
  • Keyword: Testing
  • STI Subject Categories: 42
  • Keyword: Voltage Drop
  • STI Subject Categories: 72
  • Keyword: Magnets
  • Keyword: Mice
  • Keyword: Heat Exchangers
  • Keyword: Helium
  • Keyword: Performance

Collection

  • Name: Office of Scientific & Technical Information Technical Reports
    Code: OSTI

Institution

  • Name: UNT Libraries Government Documents Department
    Code: UNTGD

Resource Type

  • Report

Format

  • Text

Identifier

  • Report No.: LBNL-3926E
  • Grant Number: DE-AC02-05CH11231
  • DOI: 10.2172/988858
  • Office of Scientific & Technical Information Report Number: 988858
  • Archival Resource Key: ark:/67531/metadc1012363