Routine Testing and Calibration Procedures for Multichannel Pulse Analyzers and Gamma-Ray Spectrometers

One of 776 items in the series: AEC research and development report available on this site.

PDF Version Also Available for Download.

Description

Report describing the results of laboratory tests with procedures for calibration of equipment and performance requirements.

Physical Description

vii, 25 p. : ill. ; 28 cm.

Creation Information

Crouch, D. F. & Heath, R. L. November 1, 1963.

Context

This report is part of the collection entitled: Technical Report Archive and Image Library and was provided by the UNT Libraries Government Documents Department to the UNT Digital Library, a digital repository hosted by the UNT Libraries. It has been viewed 427 times. More information about this report can be viewed below.

Who

People and organizations associated with either the creation of this report or its content.

Publisher

Audiences

We've identified this report as a primary source within our collections. Researchers, educators, and students may find this report useful in their work.

Provided By

UNT Libraries Government Documents Department

Serving as both a federal and a state depository library, the UNT Libraries Government Documents Department maintains millions of items in a variety of formats. The department is a member of the FDLP Content Partnerships Program and an Affiliated Archive of the National Archives.

Contact Us

What

Descriptive information to help identify this report. Follow the links below to find similar items on the Digital Library.

Titles

Description

Report describing the results of laboratory tests with procedures for calibration of equipment and performance requirements.

Physical Description

vii, 25 p. : ill. ; 28 cm.

Notes

Abstract: "Procedures for the routine calibration of gamma-ray scintillation spectrometers and multichannel pulse-height analyzers are presented. A standard laboratory scintillation spectrometer, designed for the application of recently developed computer data processing techniques, is described in detail. The performance specifications required of all electronic components in this application are discussed. In particular, the performance specifications for multi-channel pulse-height analyzers, are treated at length and simple laboratory tests are described which will determine the performance specifications of an analyzer. These tests will include live-timer accuracy, zero and ADC stability with count rate, ADC linearity, and long-term system stability."

Subjects

Library of Congress Subject Headings

Language

Item Type

Identifier

Unique identifying numbers for this report in the Digital Library or other systems.

  • OCLC: 37193148
  • SuDoc Number: Y 3.At 7:22/IDO-16923
  • Grant Number: AT(10-1)-205
  • Report No.: IDO-16923
  • Archival Resource Key: ark:/67531/metadc100240

Collections

This report is part of the following collection of related materials.

Technical Report Archive and Image Library

The Technical Report Archive & Image Library (TRAIL) identifies, acquires, catalogs, digitizes and provides unrestricted access to U.S. government agency technical reports. The mission of TRAIL is to ensure preservation, discoverability, and persistent open access to government technical publications regardless of form or format.

What responsibilities do I have when using this report?

When

Dates and time periods associated with this report.

Creation Date

  • November 1, 1963

Added to The UNT Digital Library

  • Sept. 25, 2014, 8:32 p.m.

Description Last Updated

  • March 12, 2019, 12:39 p.m.

Usage Statistics

When was this report last used?

Yesterday: 0
Past 30 days: 1
Total Uses: 427

Interact With This Report

Here are some suggestions for what to do next.

Start Reading

PDF Version Also Available for Download.

International Image Interoperability Framework

IIF Logo

We support the IIIF Presentation API

Crouch, D. F. & Heath, R. L. Routine Testing and Calibration Procedures for Multichannel Pulse Analyzers and Gamma-Ray Spectrometers, report, November 1, 1963; Idaho Falls, Idaho. (https://digital.library.unt.edu/ark:/67531/metadc100240/: accessed January 20, 2025), University of North Texas Libraries, UNT Digital Library, https://digital.library.unt.edu; crediting UNT Libraries Government Documents Department.

Back to Top of Screen